Choosing the right pressure reducer and flowmeter

In short: Two questions determine the correct part: Does the gas come from the Bottle or from the central gas supply (ZGA)? And by which Gas type Is that possible? Oxygen and compressed air have separate connections and are not interchangeable.

1. bottle or central supply

Gas source What you need In the product range
gas cylinder Pressure reducer that reduces the bottle pressure to working pressure Duostar 2, Rescue 25, Rescue 25+, MediSelect 25, Heyer
Central gas supply Quantity regulator or flow regulator, the pressure is already applied Debson flow controllers, individual flow controllers, distribution manifolds
Both alternately Pressure reducer with ZGA coupling Rescue 25+

Many pressure reducers for cylinders already have the flow regulator built in. Pressure reduction is not necessary at the central supply; there you only regulate the flow rate.

2. Gas type: Oxygen or compressed air

Each type of gas has its own connections. Therefore, we carry the same regulators in two versions, such as the Debson volume regulator and the Heyer flow regulator, each for oxygen and compressed air. When ordering, please pay attention to the information in parentheses after the product name.

  • For connection to the central supply, we offer the Y-DIN coupling ZGA for oxygen.
  • The Venturi suction device works with compressed air and uses it to create the suction.

3. Flow in stages or continuously

Stepwise regulators lock into place at fixed increments and can be read from both above and below. The range includes stepwise regulators from 0 to 25 l/min with the following increments: 0, 1, 2, 3, 4, 5, 6, 7, 9, 12, 15, and 25 l/min. Continuously variable regulators allow for intermediate values but require more precise readings.

design Flow area feature
Single flow regulator ZGA up to 15 l/min a departure, for a consumer
Distribution strip up to 15 l/min per regulator two flow controllers on one connection
Bottle pressure reducer up to 25 l/min Manometer rotatable through 360 degrees

4. Frequently Asked Questions

What is the difference between a pressure reducer and a flowmeter?
The pressure regulator reduces the high cylinder pressure to a usable working pressure. The flow meter or flow regulator then sets the flow rate in liters per minute. You need both at the gas cylinder, often combined in a single device. At the central supply, the flow regulator alone is sufficient.
Can I operate an oxygen regulator with compressed air?
No. Oxygen and compressed air have different connections to prevent exactly that from happening. That's why we offer the same regulator types in separate versions for each gas type. The gas type is indicated in parentheses after the product name.
What does the abbreviation ZGA mean?
ZGA stands for central gas supply system. The gas comes from the wall outlet and is already at working pressure. Therefore, you don't need a pressure reducer, but rather a flow regulator with a suitable coupling.
What do I need a distribution strip for?
If two consumers are to be connected to one outlet, the manifold can accommodate two flow regulators, each with a maximum flow rate of 15 l/min, at a single connection. Alternatively, you can use two individual flow regulators at separate locations.
Adjustable in steps or continuously, which is better?
Step-by-step controls click into place and are quicker and easier to read, even from the side. Continuously variable controls allow for intermediate values. Which type is suitable depends on the application and what type of thermostat is already installed in your home.
Do you also carry accessories for compressed air extraction?
Yes. The Venturi suction device generates suction from compressed air and does not require its own pump. It belongs to our product range for gas supply, not to our electric suction pumps.

5. Assignment by nib

Please tell us the type of gas, the gas source, and the required flow rate. If you are replacing an existing part, a photo of the connection will be helpful. We will then provide you with the appropriate model and price.

Telephone 02363 3663-0 · kontakt@nispel.de · Request for quotation

Last updated: August 25, 2026. The information is taken from the manufacturer's documentation for the items in our product range. The specifications of your gas supply system and the operating instructions for the appliance are binding.